6/3/2023 0 Comments Api test staphylococcusSpecies correctly identified included 94% (116 of 123) of Staphylococcus epidermidis isolates, 98% (63 of 64) of S. However, if a supplemental test was performed (most often novobiocin susceptibility), correct classification of an additional 25.7% (77) was possible, for a total of 92.7% of isolates identified to the species level. Use of the STAPH-IDENT system alone allowed correct or partly correct classification of 67% (201 of 300) of the study isolates. STAPH-IDENT is a miniaturized biochemical test strip that incorporates four synthetic chromogenic substrates, urea, arginine, and four carbohydrates and that requires only a 5-h test period. A group of 300 clinically derived isolates of coagulase-negative staphylococci were tested in parallel with the API STAPH-IDENT system (Analytab Products) and 14 conventional biochemical tests contained in Kloos and Schleifer's simplified scheme for identification of human Staphylococcus species.